Polarimetry Calibration Checklist

Switch on the lights in the observing room, and carefully install the calibration optics, then follow the checklist below (for each step, items in bold denote changes from the previous step).

X Mode Repeats Beam Blocked? LP angle Retarder?
  Rotate the LP to 0$^\circ$ (40$^\circ$)
  Pol.sum Dark 5 Yes 0$^\circ$ No
  Pol.sum 5 No 0$^\circ$ No
  Pol.sum Calib 1 No 0$^\circ$ Yes
  Pol.sum Dark 5 Yes 90$^\circ$ Yes
  Pol.sum Calib 1 No 90$^\circ$ Yes
  Pol.sum 5 No 90$^\circ$ No
  Rotate the LP to 45$^\circ$ (85$^\circ$)
  Pol.sum Dark 5 Yes 45$^\circ$ No
  Pol.sum 5 No 45$^\circ$ No
  Pol.sum Calib 1 No 45$^\circ$ Yes
  Pol.sum Dark 5 Yes 135$^\circ$ Yes
  Pol.sum Calib 1 No 135$^\circ$ Yes
  Pol.sum 5 No 135$^\circ$ No
  Rotate the LP to 180$^\circ$ (220$^\circ$)
  Pol.sum Dark 5 Yes 180$^\circ$ No
  Pol.sum 5 No 180$^\circ$ No
  Pol.sum Calib 1 No 180$^\circ$ Yes
  Pol.sum Dark 5 Yes 270$^\circ$ Yes
  Pol.sum Calib 1 No 270$^\circ$ Yes
  Pol.sum 5 No 270$^\circ$ No
  Rotate the LP to 225$^\circ$ (265$^\circ$)
  Pol.sum Dark 5 Yes 225$^\circ$ No
  Pol.sum 5 No 225$^\circ$ No
  Pol.sum Calib 1 No 225$^\circ$ Yes
  Pol.sum Dark 5 Yes 315$^\circ$ Yes
  Pol.sum Calib 1 No 315$^\circ$ Yes
  Pol.sum 5 No 315$^\circ$ No

You can now remove the calibration optics.

No. frames: ........................
Date: ........................