In order to perform the calibration it is necessary to mount the polarimetry optics over the AO and CT mirrors (near the primary focus). The mounting is very heavy, and sits on top of some very delicate (and expensive!) instrumentation, so please take great care while installing. The installed mounting can be seen in the figure below.
This calibration mounting consists of a red filter (1), a linear polarizer (LP) (2) and a removable retarder (3). During the calibration the LP is moved in steps of 45,
starting at 0
, and going to 315
(a total of 8 angles).
The angle can easily, and accurately, be changed by 90 (in one
direction) by moving the lever from position 1 to 2
(see figure 4). Making steps of 45
requires you to rotate the
angle scale (figure 3, right). Note that the angle you set is the
angle at the white tape arrow (when the lever is at position 1). The
most convenient stepping procedure needing the fewest movements of the
optical components is 0
, pos. 1-2
90
,
rotate angle scale at pos. 1
45
, pos. 1-2
135
, rotate ...
180
,
270
, 225
, 315
.
For every LP angle step you should take images in 3 camera modes:
Frame numbers and step size should be set based on your calibration analysis software and sensitivity needs. The table below outlines three possibilities:
Mode | Frames | Step Size (deg) | Expected Time |
---|---|---|---|
Long | 50 | 5 | 2 hours |
Medium 1 | 25 | 5 | 1 hour |
Medium 2 | 50 | 25 | 1 hour |
Short | 25 | 15 | 40 mins |
The retarder steps are set in a config pointed to by the POLAR_CFG field of the SOUP config file. The default, polar.cfg, has 74 angles of 5 step size. If you are in a hurry, polar_coarse.cfg can be used which has 15
steps. You are now ready to run the calibration.